[PATCH 0/3] iio: adc: ti-ads1298: Three driver cleanups

Md Shofiqul Islam posted 3 patches 2 weeks, 5 days ago
There is a newer version of this series
drivers/iio/adc/ti-ads1298.c | 11 ++++-------
1 file changed, 4 insertions(+), 7 deletions(-)
[PATCH 0/3] iio: adc: ti-ads1298: Three driver cleanups
Posted by Md Shofiqul Islam 2 weeks, 5 days ago
Three small cleanups to the TI ADS1298 medical ECG ADC driver:

1. Add missing parentheses around macro parameter in ADS1298_REG_CHnSET()
   to follow kernel macro coding style.

2. Fix a comment that incorrectly states 40ms for a 250Hz conversion
   period — the correct value is 4ms.

3. Remove the internal test signal configuration from init. The driver
   was enabling INT_TEST, TEST_AMP, and TEST_FREQ_FAST during device
   initialization, which is inappropriate for a medical ADC in normal
   operation. Only the RESERVED bit (required by the datasheet) is kept.

Md Shofiqul Islam (3):
  iio: adc: ti-ads1298: Add parentheses around macro parameter
  iio: adc: ti-ads1298: Fix incorrect timeout comment
  iio: adc: ti-ads1298: Don't enable internal test signal during init

 drivers/iio/adc/ti-ads1298.c | 11 ++++-------
 1 file changed, 4 insertions(+), 7 deletions(-)

-- 